Table 5-1. Instrumental Test Method Specifications


Table 5-1.  Instrumental Test Method Specifications



     Performance Test            

----------------------------     ---------------------------------

Analyzer Calibration Error       +- 2% of span for zero, mid-, and 

                                 high-range calibration gases



Sampling System Bias             +- 5% of span for zero and upscale

                                 calibration gases



Zero Shift                       +- 3% of span over test run period



Upscale Calibration Drift        +- 3% of span over test run period



Interference Check               +- 7% of the modified Method 6 result

                                 for each run